24-Hour Multi-Patient Use Endoscope Connectors: Letter to HealthCare Providers and Healthcare Facilities - Risk of Cross-Contamination
the recommended instructions and device design for the ERBEFLO port connector do not adequately mitigate the risks of cross-contamination for endoscopy patients. This is because the connector, tubing, and/or water bottle can become contaminated with blood, stool, or other fluids from previous patients that travel back through the endoscope channels into the connector and tubing.