UPDATE: Risk of Device Failures for Getinge’s Maquet/Datascope Cardiosave Intra-Aortic Balloon Pump (IABP) – Letter to Health Care Providers
Cardiosave IABP devices may unexpectedly shutdown or interrupt therapy (pump stops).
Cardiosave IABP devices may unexpectedly shutdown or interrupt therapy (pump stops).
The Cardiosave IABP may stop unexpectedly when internal temperatures exceed 176 degrees Fahrenheit.
Cardiosave IABPs are used to inflate and deflate intra-aortic balloons. The device may stop unexpectedly due to Autofill Failure Alarms and will be put in standby mode.
Cardiosave IABPs are used to inflate and deflate intra-aortic balloons. If they become unseated from the hospital cart, they can shut down without warning.
The Cardiosave IABP may stop unexpectedly due to a high priority Gas Gain or Gas Loss alarm. The alarm will interrupt therapy and require user intervention to correct and restart.
Wed, 30 Aug 2023 11:47:29 -0400
Dans le traitement de la spondylarthrite ankylosante évolutive.
Wed, 30 Aug 2023 09:35:05 -0400
Dans le traitement de la schizophrénie chez l'adulte.
Hamilton Medical Inc. recalls HAMILTON-C1, C2, C3, T1 Ventilators for software issues that may cause ventilators to stop without notice after approximately 91 days of cumulative use.