On July 1, 2023, Universal Meditech Inc. initiated a nationwide recall of products it manufactured between March 2021 and November 2022. The recall is due to the reasons that: i.The company is going out of business, hence would not be able to continue fulfilling any post-market responsibilities of
New Castle, Delaware, Marlex Pharmaceuticals, Inc. is voluntarily recalling one lot of Digoxin Tablets USP, 0.125mg and one lot of Digoxin Tablets USP, 0.25mg to the consumer level due to Label Mix-Up. Bottles of Digoxin Tablets, USP 0.125mg s are incorrectly labeled and contain Digoxin Tablets USP,
Cardiosave IABP devices may unexpectedly shutdown or interrupt therapy (pump stops).
The Cardiosave IABP may stop unexpectedly when internal temperatures exceed 176 degrees Fahrenheit.
Cardiosave IABPs are used to inflate and deflate intra-aortic balloons. The device may stop unexpectedly due to Autofill Failure Alarms and will be put in standby mode.
Cardiosave IABPs are used to inflate and deflate intra-aortic balloons. If they become unseated from the hospital cart, they can shut down without warning.
The Cardiosave IABP may stop unexpectedly due to a high priority Gas Gain or Gas Loss alarm. The alarm will interrupt therapy and require user intervention to correct and restart.